2D-Mode-Specific Selections For 3-Scape Imaging - Siemens ACUSON Antares Instructions For Use Manual

Ultrasound imaging system
Table of Contents

Advertisement

3 S y s t e m C o n t r o l s

2D-Mode-Specific Selections for 3-Scape Imaging

Menu
Selection
Tool Tip
(current
Rendering Mode
setting)
Plane
Cut Plane
Maps
Maps
DR
Dynamic Range
Tint
Tint
3 - 5 4
Description
Renders the 2D-mode data in the volume using the
selected rendering method:
Opacity—Smoothes image contours, creating a soft,
sculptured appearance for highlighting surface
features. Uses the selections for opacity percentage,
thresholds, and smoothing.
MinIP (Minimum Intensity Projection)—Limits the
display to minimum-value pixels by selecting pixels
with the lowest value along each projection of the
volume. May be helpful in visualizing hyper-echoic
structures and fetal skeletons.
MaxIP (Maximum Intensity Projection)—Limits the
display to maximum-value pixels by selecting pixels
with the highest value along each projection of the
volume. May be helpful in visualizing hypo-echoic
lesions and large vascular structures.
MeanIP (Mean Intensity Projection)—Limits the
display to average-value pixels by selecting pixels
with the arithmetic mean (average) value along each
projection of the volume.
Enables or disables the specified cut plane for 2D-mode
data in the volume.
Note: This selection is not available when VOI is enabled.
Selects a processing curve that assigns echo amplitudes
to gray levels.
Note: For 2D-mode, applies to slice quadrants only. The
settings J, K, and L are available for volumes acquired with
CCAI (Cadence Contrast Agent Imaging) only.
Controls the overall contrast resolution.
Applies the selected tint to the 2D-mode data.
Settings
Opacity
MinIP
MaxIP
MeanIP
Off, 1, 2
A, B, C, D, E,
F, G, H, I, J,
K, L
Transducer-
dependent
Transducer-
dependent
[ 1 ]
I N S T R U C T I O N S