Environmental Requirements; Temperature-What To Watch Out For - Agilent Technologies 85050B User's And Service Manual

7 mm calibration kit
Table of Contents

Advertisement

Specifications

Environmental Requirements

Environmental Requirements
Table 2-1
Environmental Requirements
Parameter
Temperature
a
Operating
Storage
Error-corrected range
Relative humidity
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
Table
2-1.
Avoid unnecessary handling of the devices during calibration because your
IMPORTANT
fingers are a heat source.
2-2
Limits
+20 °C to +26 °C
−40 °C to +75 °C
b
± 1 °C of measurement calibration temperature
Type tested, 0% to 95% at 40 °C, non-condensing