Troubleshooting; Symptom; Reloading Firmware - Fluke VIBXPERT II User Manual

Data collector & vibration analyzer
Hide thumbs Also See for VIBXPERT II:
Table of Contents

Advertisement

6-18
Appendix - Troubleshooting

Troubleshooting

Symptom:

VIBXPERT will not start up. The start procedure is not executed
completely, the display stops at the screen showing the LINUX
penguin.
Cause: Failure in the flash memory.
Solution: Reload the firmware.

Reloading firmware

Before reloading the firmware, ensure that the following
components are at hand or installed on the PC:
- VIBXPERT update tool version 1.47 or higher (see P. 2-37).
- Serial cable, VIB 5.430-2 and USB-Serial adapter, VIB 5.448, if
applicable.
- Ethernet cable - VIB 5.331
• Connect VIBXPERT to a laptop / desktop PC as follows:
- connect the digital interface (yellow socket) to a serial port via
cable VIB 5.430-2. If no serial interface is available, then use a
USB port and the additional USB serial adapter VIB 5.448.
- connect the communication interface (green socket) to the
VIB 5.331
VIB 5.430-2 +
VIB 5.448
network connector of the PC via Ethernet cable VIB 5.331.
• Start the VIBXPERT update tool on the PC.
• Click on the 'Start Emergency Recovery' button.
RJ 45
USB

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vib 5.310

Table of Contents