Advanced VIBCODE handling
When measurement tasks performed at a VIBCODE measurement
location are not only taken by the VIBCODE sensor but by another
vibration sensor, VIBXPERT recognizes these and assigns them to
another measurement channel.
- Enabled: Measurement tasks with the VIBCODE sensor
are performed on the VIBCODE sensor channel. The other
measurements are taken on the other channel.
- Disabled: The measurement tasks are measured on one channel,
each with the sensor that has been set up.
Disable sensor detection
This shortens the duration of the entire measurement.
· Enabled: Sensor detection is deactivated. The sensor symbol
is crossed out. If several measurements are taken of a single
measurement location, VIBXPERT only performs the sensor
detection and cable test before the first measurement.
· Disabled: Sensor detection is active. Prior to every measurement,
VIB XPERT checks whether the correct sensor is connected and if
the cable is OK. A sensor icon appears at the upper edge of the
screen next to the battery icon.
When measurements are taken with the VIBCODE sensor, sensor
detection is always active!
Disable hardware settling
This shortens the duration of the entire measurement.
- Enabled: Amplifiers do not settle before measurement starts.
- Disabled: Amplifiers settle before measurement starts.
Measurement error may occur if the amplifiers do not settle first!
When sensor detection is active (see above), the amplifiers always
settle. The settings in this field are ignored.
Triax sensor
Use the triaxial sensor in the route. The ICP sensors that were
assigned to the triax measurement location in OMNITREND are only
replaced by the triax sensor if this option is activated. Also, the
measurement tasks are grouped into the three spatial directions (X,
Y and Z) to be measured according to the following scheme:
Measure X and Y signals simultaneously on channels A and B,
then change to channel B to measure the Z signal.
Getting started - Basic settings
Sensor detection
ON
Note
Note
2-17
OFF