3-36

Recording

* Firmware module 'Recording' (VIB
5.385) must be registered.
Opening the recording
configuration window
Recording
With the 'Recording'* function, you can perform measurements
dependent on time or rotational speed. For example, this method
can be used to record measurement values under certain operating
conditions (= rpm range) or at fixed time intervals.
Availability and restrictions
Recording is available in 'Multimode' for every measurement. The
following measurements cannot be recorded:
- Coastdown curve
- Impact test
- Time waveform recorder
- Phase - Cross Channel
- All VIBCODE-measurements
Restrictions also apply to the following measurements:
- Spectrum: Averaging is not possible for rpm-controlled
recordings.
- Time waveform: The 'time synchronous' averaging type is not
permissible.
- Multi measurement tasks: rpm-controlled recording is not
possible.
- Shock pulse measurement (roller bearing condition): rpm-
controlled recording is only possible without normalization.
Configuration
• Click on 'Multimode' in the start screen.
• Select the measurement for which you want to set up recording.
• Press the MENU key and click on 'Recording' (see below).