Fluke VIBXPERT II User Manual page 35

Data collector & vibration analyzer
Hide thumbs Also See for VIBXPERT II:
Table of Contents

Advertisement

2-22
Getting started - Basic settings
Checking the measurement
path on channel B
Default sensor for vibration measure-
ment in 'Multimode'
Transducer detection and Hardware settling
In this menu, the transducer detection feature and the settling of the
amplifier stages in VIBXPERT can be switched on and off. Additional
information can be found in the 'Route Setup' section (p. 2-17).
The settling of the amplifier can only be activated when the trans-
ducer detection is switched off.
N ote
The settings in this menu apply to the 'Multimode' operating mode
only. The settings for the 'Route' and 'Machine template' modes are
made in the 'Route Setup' menu.
After the device is switched on, the amplifiers always settle prior to
the first measurement.
Sensor test
Here you can manually check the measurement path to the sensor.
For ICP sensors, use the 'ICP Sensor Test' menu; for all other sensors,
use the 'Sensor Test' menu.
• Select the channel to which the sensor is connected (A, B).
• Click on START.
VIBXPERT then checks the measurement path and displays the
result ('LineDrive', 'Open line', 'Short circuit', etc.).
Default vibration sensor (Multimode)
A default sensor can be specified for all vibration measurements in
the 'Multimode' mode. In the route / machine template, a default
sensor can only be set up via the 'Runtime setup' menu (p. 3-18) or
the Route Device Setup (p. 2-17) respectively.
• Activate the check box and select the vibration sensor.
Left:
Right: