Test 4: Light - Nellcor OxiMax SRC-MAX Technical Manual

Pulse oximetry functional tester
Table of Contents

Advertisement

SRC-MAX Operation with a Pulse Oximeter or Monitor
4. The Modulation LED indicator on the SRC-MAX will
5. The Plethysmographic waveform or blip bar on the pulse
6. Perform Test 1: BPM on page 14.
7. Perform Test 2: %SpO

Test 4: Light

The light test checks to ensure that the pulse oximeter or
monitor is functioning properly under low and high light
conditions. The pulse oximeter or monitor shall display the
correct %SpO
and 2, under low and high light conditions.
Prerequisite Procedure — Test Setup on page 12.
1. Perform Test 1: BPM on page 14.
2. Perform Test 2: %SpO
3. Press and release the Light button on the SRC-MAX.
4. The HIGH (arrow up) LED will illuminate.
5. Pleth View — The Plethysmographic waveform on the
6. Perform Test 1: BPM on page 14.
7. Perform Test 2: %SpO
16
switch from the low (arrow down) value and illuminate
the high value (arrow up).
oximeter or monitor display will increase in amplitude
within a few seconds.
and pulse rate reading, as defined in tests 1
2
pulse oximeter display may briefly disappear and then
reappear slightly altered from the previously displayed
waveform.
Blip View — The blip display will saturate momentarily
and then return to normal indications.
on page 15.
2
on page 15.
2
on page 15.
2