H2 Stationary GrainGage User's Guide
4. 2. 5 Manually Adjust a Moisture Cur ve
If desired, you can manually adjust a moisture curve using one or more grain samples.
Note: Manually adjusting the curve should be done by more experienced users. If you are unfamiliar with
moisture curves, call a HarvestMaster Field Service Engineer to walk you through the adjustment. See
8.4.2 Contact a HarvestMaster Field Service Engineer on page 93.
1. Test the grain samples in a benchtop sensor by cycling sub-samples from each larger sample 3–5
times and averaging the moisture readings. This is the Known Moisture.
2. Within the hour, take the samples out to the GrainGage. If you are calibrating outside, make sure the
samples stay in the shade.
7. Pour the grain sample into the hopper (to simulate harvest).
H2 Stationary GrainGage Calibration
3. Open Mirus.
4. Go to Setup > H2 GrainGage > Moisture Curves.
5. Select the moisture curve you want to test.
6. Tap the edit icon
.