H2 Stationary GrainGage User's Guide
H2 Stationary GrainGage Calibration
12. Pour a grain sample into the GrainGage directly onto
the moisture blade.
Note: The picture to the left shows the minimum
amount of grain needed to cover the moisture blade.
13. Enter the Known Moisture.
14. Tap Click to run a sample.
The GrainGage cycles the grain, and Mirus displays the
voltage and moisture percentage.
15. Collect the same sample and pour it into the
GrainGage again.
Tap Last Value to automatically populate the Known
Moisture with the same value as the previous sample.
16. Repeat steps 12–15 until all grain samples have been
cycled through the GrainGage.
Note: HarvestMaster recommends using three to four
samples and cycling them three times each.
If the moisture content of your samples does not cover a
range of at least 3%, Mirus displays an Insufficient Moisture
Range warning.
• Tap Ignore to continue with the current samples.
• Tap Return to add more samples with a variety of
moisture percentages.