H2 Stationary GrainGage User's Guide

4.1. 2 Check Weigh Bucket Calibration

Weigh bucket calibration is stable. It rarely changes or loses accuracy. However, because the consequences
of poor calibration can be devastating to your research data, HarvestMaster highly recommends checking
the weigh bucket calibration every year before the harvest season begins, as well as every day before
harvest. This can be done on the Weight Diagnostics screen.
H2 Stationary GrainGage Calibration
1. After calibrating the weigh bucket, tap the gear icon
below the calibration wizard to open Diagnostics.
2. Place the calibration weight diagonally across the grill.
3. The Diagnostics screen should report Weight
readings that are very close to the value stamped on
the calibration weight. If it's not within .05 lb (23 g)
HarvestMaster recommends a new weight calibration.
4. Tap the check icon
to close Diagnostics.