H2 Stationary GrainGage User's Guide
7. Pour the grain sample into the hopper (to simulate harvest).
8. Tap the Cycle button
Mirus displays the moisture, voltage, bucket weight, and temperature of the sample.
9. Cycle the sample through the GrainGage (repeat steps 7 and 8) two more times. This allows Mirus to
average the moisture readings for the sample.
10. If the moisture curve needs to be tuned to match the sample, tap Recalibrate.
The Recalibrate feature is designed to adjust previously created moisture curves to match a new grain
sample. The slope of the curve is not changed. The entire curve shifts to match the new sample. Fine tuning
of a curve is often done later in the season as additional moisture samples are available.
H2 Stationary GrainGage Calibration
.
11. Select Moisture Adjust and tap the next arrow
Note: Multiple samples can be used to adjust an
existing moisture curve.
New
sample
readings
.
55