11.8 Troubleshooting

11.8.1 Error Messages

11.8.2 Common
Problems and Solutions
Erratic Display
+OVLD or -OVLD on Display
Analog Output Incorrect
Auxiliary Function Pins Not
Operating
Sensitivity to EMI/RFI
Non-linearity at the end of
the LVDT's mechanical range
See "Error Messages" on page 151 for information relating to error messages.
Check electrical connections for continuity and the transducer's wiring code from its
Certificate of Calibration.
Make sure that the displacement on the transducer is constant.
Check the input to the (+)Signal ("+SIG") and (-)Signal ("-SIG") pins with an RMS volt-
meter in its "AC Voltage" mode. Using a voltmeter in its DC Voltage mode will always
display 0.
Indicates that the voltage across the (+)Signal ("+SIG") and (-)Signal ("-SIG") pins is
overranging or underranging the amplifier circuit. Make certain all wires are con-
nected properly, the "
CALIBRATION DATA
that a calibration has been performed per "Calibration Procedure" on page 81.
If you move the armature back to its electrical null point and you still see this mes-
sage, the (+)Excitation ("+EXC") or (-)Excitation ("-EXC") pins may be shorted to the
(+)Signal ("+SIG") or (-)Signal ("-SIG") pins.
Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see "Identifying the Output Type" on page 94.
Use the
->
DIAGNOSTICS
DAC FULL SCALE
force the Analog Output to a known output. Then, adjust your readout device, panel
meter, PLC or data acquisition system to match.
Check the output selection jumpers; see "Output Selection" on page 94.
Make sure that the
AUX1 FUNCTION
if they are set to
then they will not operate.
DISABLED
The Auxiliary Function ("AUX1" and "AUX2") pins must be connected to pin 10, not pin
8, to activate them.
To obtain maximum immunity to electromagnetic or radio frequency interference,
make certain that the shields of the transducer cables are connected to the "cable
shield connection screw" on the rear panel of the instrument. See "Connection of
Four- or Five-wire AC/AC-LVDT" on page 80, "External Arrangement of AC powered
SC1000 and SC2000" on page 27 and "External Arrangement of Model SC3004" on
page 33.
Most LVDTs have a usable measuring range that is smaller than their mechanical
range. See "Electrical Null and Transducer Mounting" on page 93 for information on
establishing the LVDT's electrical null point to insure that the LVDT will be operated in
its usable measuring range.
->
" menu item is set correctly, and
FULL SCALE VRMS
and
->
DIAGNOSTICS
DAC ZERO SCALE
or
SETUP menu items are set correctly;
AUX2 FUNCTION
SC Series Instruction Manual
menu items to
page 95