10.4 Specifications

page 62
008-0608-00
TRANSDUCER INPUT
Transducer type
Excitation Voltage
Transducer full-scale output
Amplifier Gain Selection
Calibration Type
Differential Input Voltage
A/D Converter
Low-pass filter
Resolution and Frequency Response
INSTRUMENT-ONLY ACCURACY
(Sense wires used;Frequency Response setting<=16Hz;
Linearity, repeatability & hysteresis)
mV/V Calibration
Known -Load Calibration
AUXILLIARY INPUTS
Response Time
Field-Selectable Functions
ANALOG OUTPUT
Output voltage range
Output current range (optional current
output channels)
Frequency Response
full-bridge, strain gage
short circuit protected with sensing
.5 to 11.0 mV/V @ 5V excitation
.5 to 5.5 mV/V @ 10 V excitation
2-, 3- or 5- point known load
24-bit Sigma-Delta
digital, 24-tap FIR
see "FREQ. RESPONSE Menu Item" on
Quantity
Type
momentary contact closure
peak/valley clear, peak/valley hold,
5, ±5, 10 or ±10 VDC (field selectable)
Source
any channel's track, peak or valley value
Isolation
Resolution
same as input when driven by the same
channel's tracking data
5 or 10 VDC,
automatic
shunt, mV/V or
+/- 55mV (max.)
page 63
±0.1% F.S.
±0.01% F.S.
2
< 5ms
tare on, tare off,
track hold
4-20 mA
500V
13 bits