Display Setup Sub-Menu - Honeywell SC1000 Owner's Manual

Sensotec sensors signal conditioning, self-calibrating digital indicators
Table of Contents

Advertisement

10.5.2
DISPLAY SETUP
Sub-Menu
Menu Item
DISPLAY. DIGITS
Menu Item
DISPLAY. DECPT
Menu Item
DISPLAY. COUNT-BY
Menu Item
DISPLAY. UNITS
Menu Item
DISPLAY. AVERAGE
This menu controls how data values are displayed by the channel and transmitted
via serial communications.
Selects the number of digits displayed by the channel. The choices are:
"
" displays both positive and negative numbers with five full dig-
5 DIGIT-BIPOLAR
its (±99999).
"
" displays positive numbers with six full digits (999999) and
6 DIGIT-UNIPLOAR
negative numbers with five full digits (-99999).
"
displays positive number with seven full digits (9999999)
7 DIGIT UNIPOLAR"
and negative numbers with six digits (-999999).
This menu item may be automatically updated by a transducer's Signature Mod-
ule.
Selects the decimal point location on the channel's display and serial communica-
tions output. Use the [UP] and [DOWN] buttons to move the decimal point to the
right and left.
This menu item may be automatically updated by a transducer's Signature Mod-
ule.
Determines by what value the display increments or decrements. This will make
the display less sensitive to a noisy signal. The choices are:
"
"
00001
"
"
00002
"
"
00005
"
"
00010
"
"
00020
"
"
00050
"
"
00100
"
"
00200
This menu item may be automatically updated by a transducer's Signature Mod-
ule.
Specifies the four character label that is displayed to the right of the channel's val-
ues.
NOTICE
This menu item doesn't change the mathematical scaling of the channel's val-
ues; that can be changed by altering the "
menu item and then performing a re-calibration.
When a character position is flashing press the [UP] or [DOWN] button to change
the character. Press [ENTER] to advance to the next character.
This menu item may be automatically updated by a transducer's Signature Mod-
ule.
Controls the speed with which the channel's display values will update. Display
averaging does not affect the channel's analog output or its peak/valley detection,
which will proceed at the speed selected by the "
item. The choices are:
"
" means that the display will update four times each second. The channel's
ON
values will be averaged for 1/4 second, then displayed.
"
" means that the channel's display will update as quickly as possible.
OFF
->
CALIBRATION DATA
FULL SCALE VALUE
->
OPERATION
FREQ. RESPONSE
SC Series Instruction Manual
"
" menu
page 65