15.4 Channel Menu

Menu Item
LEFT CHANNEL
Menu Item
LEFT SOURCE
Menu Item
RIGHT CHANNEL
Menu Item
RIGHT SOURCE
Menu Item
LEAVE MENU

15.5 Troubleshooting

Front Panel Buttons not
Operating in RUN Mode
page 140
008-0608-00
Detailed instructions on operating the instrument in the SETUP Menu mode can be
found in "SETUP Menu mode" on page 23. A diagram of all menus is located in
"Setup Menu Reference" on page 157.
This specifies the channel that is displayed on the left half of the split display.
This lets you designate the data source of the channel monitored by the left half of the
split display. Each channel has three data sources: the live tracking value (TRACK),
its highest value (PEAK), and its lowest value (VALLEY).
The options for this menu item are:
"
" means the live tracking value of the channel.
TRACK
"
" means the highest value of the channel since the peak detector was last
PEAK
cleared.
"
" means the lowest value of the channel since the valley detector was last
VALLEY
cleared.
This specifies the channel that is displayed on the right half of the split display.
This lets you designate the data source of the channel monitored by the right half of
the split display.
Each channel has three data sources: the live tracking value
(TRACK), its highest value (PEAK), and its lowest value (VALLEY).
The options for this menu item are:
"
" means the live tracking value of the channel.
TRACK
"
" means the highest value of the channel since the peak detector was last
PEAK
cleared.
"
" means the lowest value of the channel since the valley detector was last
VALLEY
cleared.
When the menu item is displayed, press [ENTER] to leave this menu. Press [EXIT] to
return to the RUN mode.
When a Split Display Virtual channel is displayed the [VALUE], [CLEAR] and [TARE]
buttons have no effect. To change one of the values shown on a Split Display Virtual
channel, use the its SETUP menu. To clear peak/valley values or tare an Input chan-
nel, use the [CHANNEL] button to select an Input channel and then use the [CLEAR]
or [TARE] buttons.