11.5.6
Menu
CALIBRATE
Item
This menu item performs a calibration according to what was entered in the
and
BRATION TYPE
CALIBRATION DATA
NOTICE
Before performing a calibration,
the transducer must be connected to the instrument, the
be selected (see "CALIBRATION TYPE Menu Item" on page 87) and the
must be entered (see "CALIBRATION DATA Sub-Menu" on page 88).
TION DATA
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
If the
is 2-Point Known Displacement Calibration...
CALIBRATION TYPE
•
The display will read
DOING 2POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
•
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
•
Upon completion, the display will indicate
the RUN mode.
If the
is 3-Point Known Displacement Calibration...
CALIBRATION TYPE
•
The display will read
DOING 3POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
•
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
•
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
•
Upon completion, the display will indicate
the RUN mode.
If the
is 5-Point Known Displacement Calibration...
CALIBRATION TYPE
•
The display will read
DOING 5POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
•
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
•
The display will read
WORKING
menu items.
, and prompt you to
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/2
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00500 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00250 UNIT
and
KNOWN POINT 2/5
DISPLAY. UNITS
, then
APPLY 00500 UNIT
SC Series Instruction Manual
CALI-
must
CALIBRATION TYPE
CALIBRA-
APPLY 00000 UNIT
and
KNOWN POINT 1/2
DISPLAY.
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/3
DISPLAY.
(where 00500 and UNIT are
). When you have
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/5
DISPLAY.
(where 00250 and UNIT are
). When you have
(where 00500 and UNIT are
page 89