Calibrate Menu Item - Honeywell SC1000 Owner's Manual

Sensotec sensors signal conditioning, self-calibrating digital indicators
Table of Contents

Advertisement

11.5.6
Menu
CALIBRATE
Item
This menu item performs a calibration according to what was entered in the
and
BRATION TYPE
CALIBRATION DATA
NOTICE
Before performing a calibration,
the transducer must be connected to the instrument, the
be selected (see "CALIBRATION TYPE Menu Item" on page 87) and the
must be entered (see "CALIBRATION DATA Sub-Menu" on page 88).
TION DATA
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
If the
is 2-Point Known Displacement Calibration...
CALIBRATION TYPE
The display will read
DOING 2POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
Upon completion, the display will indicate
the RUN mode.
If the
is 3-Point Known Displacement Calibration...
CALIBRATION TYPE
The display will read
DOING 3POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
Upon completion, the display will indicate
the RUN mode.
If the
is 5-Point Known Displacement Calibration...
CALIBRATION TYPE
The display will read
DOING 5POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this displacement to the transducer, press
UNITS
[ENTER].
The display will read
WORKING
previously entered as
applied this displacement to the transducer, press [ENTER].
The display will read
WORKING
menu items.
, and prompt you to
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/2
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00500 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00250 UNIT
and
KNOWN POINT 2/5
DISPLAY. UNITS
, then
APPLY 00500 UNIT
SC Series Instruction Manual
CALI-
must
CALIBRATION TYPE
CALIBRA-
APPLY 00000 UNIT
and
KNOWN POINT 1/2
DISPLAY.
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/3
DISPLAY.
(where 00500 and UNIT are
). When you have
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/5
DISPLAY.
(where 00250 and UNIT are
). When you have
(where 00500 and UNIT are
page 89