14.7 Troubleshooting

14.7.1 Error Messages

14.7.2 Common
Problems and Solutions
Analog Output Incorrect
Sensitivity to EMI/RFI
page 136
008-0608-00
See "Error Messages" on page 151 for information relating to error messages.
Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see "Identifying the Output Type" on page 135.
Use the
->
DIAGNOSTICS
DAC FULL SCALE
force the Analog Output to a known output. Then, adjust your readout device, panel
meter, PLC or data acquisition system to match.
Check the output selection jumpers; see "Output Selection" on page 135.
Check the "
->
DAC SETUP
DAC. ZERO-SCALE
To obtain maximum immunity to electromagnetic or radio frequency interference,
make certain that the shield of the cable is connected to ground/earth only at the read-
out device. The shield of the cable should not be connected to the instrument.
and
->
DIAGNOSTICS
DAC ZERO SCALE
" and "
->
DAC SETUP
DAC FULL-SCALE
menu items to
" menu items.

Advertisement