10.5.6
Menu
CALIBRATE
Item
This menu item performs a calibration according to what was entered in the
and
BRATION TYPE
CALIBRATION DATA
NOTICE
Before performing a calibration, the transducer must be connected to the
instrument, the
CALIBRATION TYPE
Menu Item" on page 67), and the
BRATION DATA Sub-Menu" on page 69).
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
If the
is Shunt Calibration...
CALIBRATION TYPE
•
The display will read
DOING SHUNT CAL
00000 and UNIT are previously entered as
S. When you have applied this load to the transducer, press [ENTER].
UNIT
•
The display will then read
pletion, the display will indicate
mode.
If the
is mV/V Calibration...
CALIBRATION TYPE
•
The display will read
DOING MV/V CAL
00000 and UNIT are previously entered as
). When you have applied this load to the transducer, press [ENTER].
UNITS
•
The display will then read
pletion, the display will indicate
mode.
If the
is 2-Point Known Load Calibration...
CALIBRATION TYPE
•
The display will read
DOING 2POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this load to the transducer, press [ENTER].
UNITS
•
The display will read
WORKING
previously entered as
applied this load to the transducer, press [ENTER].
•
Upon completion, the display will indicate
the RUN mode.
If the
is 3-Point Known Load Calibration...
CALIBRATION TYPE
•
The display will read
DOING 3POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this load to the transducer, press [ENTER].
UNITS
•
The display will read
WORKING
previously entered as
applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
previously entered as
menu items.
must be selected (see "CALIBRATION TYPE
must be entered (see "CALI-
CALIBRATION DATA
, and prompt you to
ZERO SCALE VALUE
and calibration will take place. Upon com-
WORKING
and the instrument will return to the RUN
DONE
, and prompt you to
ZERO SCALE VALUE
and calibration will take place. Upon com-
WORKING
and the instrument will return to the RUN
DONE
, and prompt you to
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/2
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00500 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
, then
APPLY 01000 UNIT
and
KNOWN POINT 3/3
DISPLAY. UNITS
SC Series Instruction Manual
CALI-
(where
APPLY 00000 UNIT
and
DISPLAY.
(where
APPLY 00000 UNIT
and
DISPLAY.
APPLY 00000 UNIT
and
KNOWN POINT 1/2
DISPLAY.
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/3
DISPLAY.
(where 00500 and UNIT are
). When you have
(where 01000 and UNIT are
). When you have
page 71