Preliminary work

Removing the swivel arm cover

Fig. 3
Removing the swivel arm cover
Remove object table (1).
Remove face shield (2) by sliding it off its holder.
Remove front cover (3) by removing two screws (4).
Remove side covers (5) by removing twelve screws (6) and disconnecting the two
connectors X807, connected to tube head control panel (7), and the two ground cables.
Remove the shaft covers (8) by removing ten screws (9) and disconnecting the two
connectors X808, connected to control panel (10)inside the shaft cover.
CAUTION
Siemens AG
Medical Engineering
5
4
6
3
2
1
Be careful not to damage connectors X807 and X808 when remov-
ing the covers.
Register 5
RXB7-230.091.05
Rev. 03
7
10
Page 3 of 8
07. 98
TD SD 24
8
9
MAMMOMAT 300 / 3000
2 - 3
2