Preliminary work
Removing the swivel arm cover
Fig. 3
Removing the swivel arm cover
•
Remove object table (1).
•
Remove face shield (2) by sliding it off its holder.
•
Remove front cover (3) by removing two screws (4).
•
Remove side covers (5) by removing twelve screws (6) and disconnecting the two
connectors X807, connected to tube head control panel (7), and the two ground cables.
•
Remove the shaft covers (8) by removing ten screws (9) and disconnecting the two
connectors X808, connected to control panel (10)inside the shaft cover.
CAUTION
Siemens AG
Medical Engineering
5
4
6
3
2
1
Be careful not to damage connectors X807 and X808 when remov-
ing the covers.