3 - 14
3
Fig. 15

Removing the cable harness at the generator

Remove the front cover (3/Fig.15) of the generator.
Remove the following connectors at the generator according to the installation
instructions:
High voltage cable.
X9 Rotating anode.
X14 Power supply for the stand.
X8 Filament.
X1 Serial interface.
X10, X11 Iontomat cable.
Protective conductor.
Pull back all cables that lead from the generator to the base of the stand.
MAMMOMAT 300 / 3000
Register 5
RXB7-230.091.05
Rev. 03
Removing de cable harness
2
1
Page 14 of 16
07. 98
TD SD 24
3
Siemens AG
Medical Engineering