3 - 14
3
Fig. 15
Removing the cable harness at the generator
•
Remove the front cover (3/Fig.15) of the generator.
Remove the following connectors at the generator according to the installation
instructions:
•
High voltage cable.
•
X9 Rotating anode.
•
X14 Power supply for the stand.
•
X8 Filament.
•
X1 Serial interface.
•
X10, X11 Iontomat cable.
•
Protective conductor.
Pull back all cables that lead from the generator to the base of the stand.
MAMMOMAT 300 / 3000
Register 5
RXB7-230.091.05
Rev. 03
Removing de cable harness
2
1
Page 14 of 16
07. 98
TD SD 24
3
Siemens AG
Medical Engineering