MAMMOMAT 3000/ 3000CE only - Removing the corrugated hose

Fig. 7
Remove the stereo switch (1).
Remove the 4 screws (2) and remove the lead-through ring (4).
Remove the hose clamp (3).
NOTICE
Pull the lead-through ring off the cable harness.
MAMMOMAT 300 / 3000
2
1
for MAMMOMAT 3000 CE:
- The braided grounding wire should remain attached to the swivel
arm.
- Unfold the shielding and remove the semicircular cable clamp.
Register 5
Removing de cable harness
3
4
RXB7-230.091.05
Rev. 03
07. 98
Page 6 of 16
TD SD 24
Medical Engineering