Mammomat 3000/ Ce - Siemens MAMMOMAT 300 Repair Instructions

Hide thumbs Also See for MAMMOMAT 300:
Table of Contents

Advertisement

4 - 10

Mammomat 3000/ CE

X896
X877
X878
X894
4 stop button
wires
X895
4 ground
wires
X896 (X-ray tube) G
Wire
Fig. 12
Divide the cable harness and insert it. Route the cable harness so that is is free of coils
and kinks.
Secure the cable with cable ties.
Reinsert wire 2 into connector X860.
Insert all cables as shown in the Fig. and connect as many plugs as possible at this point.
MAMMOMAT 300 / 3000
X893
H3
G
0
Register 5
Installing the cable harness
H1M
H2M
H1W
1
2
3
RXB7-230.091.05
Rev. 03
07. 98
X897
X870
X858
X874
X856
X860
X855
High voltage cable
H2W
no designation
Page 10 of 14
TD SD 24
Medical Engineering