3 - 2
Remove the connectors at the MAMMOMAT 3000/3000 CE swivel
arm
1 ground cable
4 emergency
stop button wires
X893
LEFT SIDE
Fig. 2
Remove all connectors in accordance with and the following list (including the protective
conductors).
The cables highlighted in gray have already been removed.
Eject wire #2 from connector X860.
Remove the four wires from the STOP switch.
Remove the wire with connector X893.
MAMMOMAT 300 / 3000
X896
X895
X81
Register 5
Removing de cable harness
1 ground cable
X894
X878
X877
X897
RXB7-230.091.05
Rev. 03
07. 98
X858
X870
X856
X855
X860
RIGHT SIDE
Page 2 of 16
TD SD 24