Remove the connectors at the MAMMOMAT 300/300 CE swivel
arm
4 emergency
stop button
cables (backside)
Fig. 1
Remove all connectors in accordance with Fig. 1 and the list which follows (including the
protective conductors).
The cables highlighted in gray have already been removed.
Remove the four wires from the STOP switch.
Siemens AG
Medical Engineering
X896
LEFT SIDE
Register 5
RXB7-230.091.05
Rev. 03
X895
X894
Page 1 of 16
07. 98
TD SD 24
3 - 1
X897
X855
X856
X860
X870
RIGHT SIDE
MAMMOMAT 300 / 3000