3 - 12
MAMMOMAT 300 CE and MAMMOMAT 3000 CE only
Remove the screws from the connection device (1/Fig. 12) on the base of the stand and
disconnect the following connectors:
•
X885 connection cable for the stereo evaluation unit.
•
Connector X885.
•
Wire to line filter Z801.
•
Protective conductor.
•
Wire on connector X803 (back).
•
Strain relief for all wires to the generator.
Fig. 12
MAMMOMAT 300 / 3000
1
Register 5
RXB7-230.091.05
Rev. 03
Removing de cable harness
Page 12 of 16
07. 98
TD SD 24
3
Siemens AG
Medical Engineering