MAMMOMAT 3000/ CE

1 ground cable
4 emergency
stop button wires
X893
LEFT SIDE
Fig. 10
Place the lead-through ring over the corrugated hose.
Attach the lead-through ring with the screws.
NOTICE
Install the stereo switch with the screws.
MAMMOMAT 300 / 3000
X896
X895
X81
Make sure that the shielding is folded over the lead-through ring
and that the semicircular cable clamp is in position. Also make
sure that the braided grounding wire is between the hose clamp
and the shielding.
Register 5
Installing the cable harness
1 ground cable
X894
X878
X877
X897
RXB7-230.091.05
Rev. 03
07. 98
X858
X870
X856
X855
X860
RIGHT SIDE
Page 8 of 14