Removing de cable harness
Mark the position of the cable (Position with reference to the stand, position of the cable
ties).
Remove the component parts of the high voltage cable plug.
Cut the cable ties from the cable harnesses being removed.
Remove the cable from the swivel arm.

MAMMOMAT 300/300 CE only - Removing the corrugated hose

Fig. 5
Fig. 6
Remove the 4 screws (1);
the table support can then be lowered by approximately 1 cm.
Remove the clamp (2).
NOTICE
Siemens AG
Medical Engineering
for MAMMOMAT 300 CE:
- The braided grounding wire should remain attached to the swivel
arm.
- Unfold the shielding and remove the semicircular cable clamp.
Register 5