Removing de cable harness
•
Mark the position of the cable (Position with reference to the stand, position of the cable
ties).
•
Remove the component parts of the high voltage cable plug.
•
Cut the cable ties from the cable harnesses being removed.
•
Remove the cable from the swivel arm.
MAMMOMAT 300/300 CE only - Removing the corrugated hose
Fig. 5
Fig. 6
•
Remove the 4 screws (1);
the table support can then be lowered by approximately 1 cm.
•
Remove the clamp (2).
NOTICE
Siemens AG
Medical Engineering
for MAMMOMAT 300 CE:
- The braided grounding wire should remain attached to the swivel
arm.
- Unfold the shielding and remove the semicircular cable clamp.
Register 5
RXB7-230.091.05
Rev. 03
R803
2
Page 5 of 16
07. 98
TD SD 24
3 - 5
1
MAMMOMAT 300 / 3000
3