Removing The Cable Harness From The Stand - Siemens MAMMOMAT 300 Repair Instructions

Hide thumbs Also See for MAMMOMAT 300:
Table of Contents

Advertisement

Removing de cable harness

Removing the cable harness from the stand

Fig. 13
Fig. 14
Loosen the hose clamp (1/Fig.13) for the corrugated hose on the lifting carriage.f
NOTICE
Siemens AG
Medical Engineering
1
for MAMMOMAT 300 CE and MAMMOMAT 3000 CE
The braided grounding wire remains attached to the lifting carriage.
Remove the angle bracket (1/Fig.14) for the corrugated hose.
Pull the corrugated hose (2/Fig.14) into the lifting carriage.
Register 5
RXB7-230.091.05
Rev. 03
2
Page 13 of 16
07. 98
TD SD 24
3 - 13
1
MAMMOMAT 300 / 3000
3