Removing de cable harness
Removing the connectors on the left side of the stand
•
X887 Lift motor.
•
X888 and X826 Swivel motor.
•
X861 Angle detector board.
•
X898 Rotation brake.
•
Protective conductor.
•
Microswitches: S882, S881, S 883, S884.
NOTICE
.
1
1
Fig. 11
Siemens AG
Medical Engineering
For some stand versions, the swivel motor must be removed
(1/Fig.11):
- Insert the transport screw for the rotation counterbalance, refer to
(1/Fig. 11).
- Remove the 3 screws show in (2/Fig. 11) and remove the swivel
motor (3/Fig. 11).
Register 5
RXB7-230.091.05
Rev. 03
3
2
Page 11 of 16
07. 98
TD SD 24
3 - 11
3
MAMMOMAT 300 / 3000