Installing the cable harness
Apply some silicon oil to the high voltage plug. Connect the plug to the tube
(see p3-4, 3-5).
Route the cables in the swivel arm as shown in Fig. 12.
Install connectors X891 and X893 as shown in Fig. 13.
To check for correct positioning, insert the table (Stereotactic biopsy unit, when used
(Wing 1)).
Make sure the connector(s) dock properly!
Connect the STOP switch.
Fig. 13

Installing X891 and X893

Yellow (YE)
Fig. 14

Connecting the STOP switch

Fig. 15
Siemens AG
Medical Engineering
Blue (BU)
STOP switch connections
white (WH)
blue (BU)
black (BK)
yellow (YE)
Register 5
Connection no.
RXB7-230.091.05
Rev. 03
07. 98
White (WH)
Black (BK)
11
12
21
22
Page 11 of 14
MAMMOMAT 300 / 3000
TD SD 24
4 - 11
4