3.1.2 Removing the Front cover.

3.1.3 Removing the X-ray source assembly.

Turn off the Multimobil 10,disconnect the mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to transport position. Lock the Multimobil 10 by
applying the brake.
Make sure that the X-ray source assembly is properly secured with the transport
safety lock. It should not be possible to move the X-ray source assembly
upwards.This is very important for safety reasons.
Siemens Ltd. Med Division India
Service Instructions
Turn off the Multimobil 10,disconnect the
mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to
transport position. Lock the Multimobil 10
by applying the brake.
Arrange some kind of soft bedding to place
the front cover on .
Unscrew the four ornamental screws with
2.5mm allen key.
Grasp the cassette box and slowly pull it
outside.
Place the front cover on a soft bedding.
Version 4.0
Service Manual
Page 17 of 48