AGFA ADC Solo User Manual page 53

Hide thumbs Also See for ADC Solo:
Table of Contents

Advertisement

Status field:
MESSAGE 1
CASSETTE NOT
IDENTIFIED
24 X 30 CM
CALIBRATION
MISSING
SERVICE MODE
CASSETTE SLOT
BLOCKED
IMAGE-QUEUE FULL
UNKNOWN
DESTINATION
[PPNAME]
RIGHT SIDE PANEL
NOT CLOSED
UNKNOWN IP-TYPE
EMERGENCY DATA
NOT DEFINED
2301D EN 20040224
LOCKED
Message 2
Press
, remove
and identify
Press
to accept
or
Please wait
Remove cassette,
press
Check queue
Press
, remove
cassette and identify
Close right side
panel
Press
, remove
cassette, call
Service.
Emergency keys
disabled, press
Action
• Press Confirm key.
• Remove cassette.
• Identify cassette.
• Re-insert cassette.
• Press Confirm key to treat
24 x 30 cm image plate with-
out calibration or press Can-
cel key to treat cassettes with
other formats.
• Contact your local service
organization.
Wait.
• Remove cassette.
• Remove obstructing objects.
• Press Confirm key.
• Refer to
'Consulting the
image transmission queue
('Queue management')'
page
48
of the Reference
manual.
• Check that the ADC Solo is
not off line (Refer to
play'
on page 15).
• Press Confirm key.
• Remove cassette.
• Identify cassette.
• Re-insert cassette.
• Check the configuration of
the system.
Close the right side panel.
• Press Confirm key.
• Remove cassette.
• Contact your local service
organization.
• Press Confirm key
• Contact your local service
organization.
on
'The dis-
53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents