Reading an image plate in the dedicated configuration
In the dedicated configuration, the ADC Solo can digitize the image plate while
you are entering the identification data via the ID Station.
To read an image plate:
1
Check that the ADC Solo is ready for operation:
• the ADC Solo must display the operator main screen with 'Ready' status, e.g.:
VIPS not ready
ADC SOLO
• the status indicator at the top of the ADC Solo must be green and be lit constantly.
The ADC Solo is operational if the status field equals 'READY', even if status
messages of the destination are shown (e.g. 'VIPS not ready').
2
Press the appropriate emergency button at the front of
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
The button which you have pressed will be lit and the display will read:
or press
ADC SOLO
28
For digitizing images of the trunk.
For digitizing images of the limbs.
IDENTIFY AT ID STATION
to use emergency data
READY
BUSY
DIRECT ID
Status field
the
ADC Solo.