AGFA ADC Solo User Manual page 32

Hide thumbs Also See for ADC Solo:
Table of Contents

Advertisement

3
Press the Emergency key on the keypad.
The display will read:
Next cassette gets emergency status
ADC SOLO
The emergency status will only be assigned to the first image plate which
you insert into the ADC Solo cassette slot after pressing the Emergency
key.
4
Insert the cassette containing the emergency image plate into the cassette slot
as shown below. Make sure to insert the cassette with the hinge [1] at the top
and the locking mechanism [2] at the bottom.
When the ADC Solo has treated the emergency image plate, it displays the operator
main screen. The image processing station will give the emergency image priority
over the other images in the image processing queue.
If you do not enter a cassette within 1 minute after pressing the Emergency
key or if you enter a cassette without ID data, the ADC Solo will quit the
emergency function and return to the operator main screen.
5
Remove the cassette from the cassette slot.
32
WARNING
1
2
READY
EMERGENCY
2301D EN 20040224