AGFA ADC Solo User Manual page 33

Hide thumbs Also See for ADC Solo:
Table of Contents

Advertisement

Reading emergency image plates without ID data
To read an emergency image plate without ID data:
1
Check that the ADC Solo is ready for operation:
• the ADC Solo must display the operator main screen with 'Ready' status, e.g.:
ADC SOLO
• the status indicator at the top of the ADC Solo must be green and be lit constantly.
The ADC Solo is operational if the status field equals 'READY', even if status
messages of the destination are shown (e.g. 'VIPS not ready').
2
Press the appropriate
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
For digitizing unidentified emergency images of the trunk.
For digitizing unidentified emergency images of the limbs.
The button which you have pressed will be lit and the display will read:
Please enter unidentified cassette
ADC SOLO
The emergency status will only be assigned to the first image plate which
you insert in the ADC Solo cassette slot after pressing the emergency but-
ton.
2301D EN 20040224
emergency button at the front of the
EMERG. BUTTON
WARNING
READY
READY
Status field
ADC Solo.
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents