4
Enter the identification data via the ID Station.
For detailed information, refer to the User manual of the ADC ID Software.
As soon as you have entered the identification data, the ADC Solo displays:
Miller
ADC SOLO
If fast preview is enabled, the image data are sent to the preview station as soon as
you have entered the identification data. The image data are sent in blocks of typical
100 lines. For more information, refer to the User manual of the ADC Preview Soft-
ware.
As soon as the ADC Solo has digitized the entire image plate and you have entered
the identification data:
• the ADC Solo erases the image plate and re-inserts it into the cassette;
• the ADC Solo gives the cassette ID data the status 'erased'.
• the cassette is returned to the cassette slot;
• the digital image data is sent to the image processing station ('destination').
When the ADC Solo has treated the cassette, it displays the operator main screen.
5
Remove the cassette from the cassette slot.
When the ADC Solo returns the cassette, it is ready to be re-used immedi-
ately. However, if you leave it for more than 3 days before re-using it, you
must re-erase it first. Refer to
30
BUSY
Chest AP
'Re-erasing an image plate'
on page 38.