Special Calibration Requirements For Dual Range Or Auto Range; References - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

EPA Protocol Calibration
Follow the procedures described in section 7.2 for calibrating the zero points.
For each mid point:
Wait until
STABIL falls
below 0.5 ppb.
This may take
several minutes.
8.6. Special Calibration Requirements for Dual Range or
Auto Range
If Dual Range or Auto Range is selected, then it should be calibrated for both Range1 and
Range2 separately.
For zero and span point calibration, follow the procedure described in Section 7.2. Repeat
the procedure for both the HIGH and LOW Ranges

8.7. References

1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50,
Appendix A, Section 10.3.
152
SAMPLE*
RANGE = 500.000 PPB
< TST
>
CAL
TST
SAMPLE
STABIL=X.XXX PPB
< TST TST >
CAL
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE
STABIL=X.XXX PPB
< TST TST > CAL
CALZ
CALS
RANGE = 500.0 PPB
SPAN CAL M
< TST TST >
ZERO SPAN
CONC
Model 100E Instruction Manual
SO2 =X.XXX
Set the Display to show the
STABIL test function.
This function calculates the
SETUP
stability of the NO/NO
SO2 =X.XXX
SETUP
SO2 =X.XXX
SETUP
Record the
SO
\reading as
2
displayed on the
SO
X.XXX
2
instrument's front
panel
EXIT
Press EXIT to
Return to the
Main SAMPLE
Display
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
045150102 Rev XB1
x
measurement
ACTION:
port