Warm-Up; Figure 3-9: Front Panel Layout; Table 3-5: Nist-Srm's Available For Traceability Of So Table 3-6: Front Panel Display During System Warm-Up - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
The analyzer should automatically switch to SAMPLE mode after completing the boot-up
sequence and start monitoring SO

3.2.2. Warm-Up

The M100E requires about 60 minutes warm-up time before reliable SO
can be taken. During that time, various portions of the instrument's front panel will
behave as follows. See Figure 3-9 for locations.
Table 3-6:
Name
Color
Concentration
Field
Mode Field
STATUS LED's
Sample
Green
Cal
Yellow
Fault
FASTENER
MODE FIELD
ADVANCED POLLUTION INSTRUMENTATION, INC.
KEYBOARD
045150102 Rev XB1
gas.
2
Front Panel Display During System Warm-Up
Behavior
N/A
Displays current,
compensated SO
Concentration
N/A
Displays blinking
"SAMPLE"
On
Off
Red
Blinking
KEY DEFINITIONS
MESSAGE FIELD
SAMPLE A
<TST
TST>
CAL
UV FLUORESCENCE SO2 ANALYZER - MODEL 100E
Figure 3-9:
N/A
2
Instrument is in sample mode but is still in the
process of warming up.
Unit is operating in sample mode, front panel
display is being updated.
The instrument's calibration is not enabled.
The analyzer is warming up and hence out of
specification for a fault-free reading. various warning
messages will appear.
CONCENTRATION FIELD
RANGE = 500.000 PPB
Front Panel Layout
Getting Started
measurements
2
Significance
STATUS LED's
SAMPLE
SO2 = 40.0
CAL
SETUP
FAULT
POWER
FASTENER
ON / OFF SWITCH
37