Advertisement

Performing User Calibration
5� Setting an ID
Names can be applied to calibration channels.
Memo
By default, the channel number is set as the name.
Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key�
The setting screen is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[User Calib], and then press the [RIGHT]
or [ENTER] key�
3
Press the [UP] or [DOWN] key to select
[Set ID], and then press the [RIGHT] or
[ENTER] key�
4
The screen for selecting a probe is
displayed� Press the [UP] or [DOWN] key
to select a probe and then press the
[RIGHT] or [ENTER] key�
5
The screen for selecting a calibration
channel is displayed� Press the [UP] or
[DOWN] key to select a channel�
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
User calib > Set ID > Select CH
01
CH00
CH01
CH02
CH03
CH04
CH05
CH06
CH07
CH08
CH09
78
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
C o p y   a n d   R e g i s t e r
E d i t   C o r r e c t i o n   C o e f
S e t I D
D e l e t e   C a l i b   D a t a
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
C o p y   a n d   R e g i s t e r
E d i t   C o r r e c t i o n   C o e f
S e t I D
D e l e t e   C a l i b   D a t a
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
01
C o p y   a n d   R e g i s t e r
02
E d i t   C o r r e c t i o n   C o e f
03
S e t I D
04
D e l e t e   C a l i b   D a t a
05
06
07
08
09
10