Konica Minolta CA-410 Instruction Manual page 20

Display color analyzer, probe+ data processor
Hide thumbs Also See for CA-410:
Table of Contents

Advertisement

Probes
Trigger timing diagram
Trigger IN
Transition to trigger mode
Trigger OUT
• To input trigger signals or external synchronization signals to the probe, input the signals to the BNC plug of an optional
conversion cable IF-A29 or an optional BNC conversion cable IF-A35.
(Luminance/chromaticity measurement: 0.5 to 240 Hz)
(Flicker measurement: 0.5 to 130 Hz)
For connection with a CA-DP40 data processor, you can use a mini DIN cable or a USB cable.
• For RS-232C connection, connect the mini DIN plug of the RS mini DIN cable IF-A30 (or IF-A31 or IF-A32) for the data
processor to the RS-232C connector of the probe and the probe RS-232C connector of the CA-DP40 data processor.
Connection via a mini DIN cable (using IF-A30/31/32)
• For USB connection, connect the USB mini B plug of the supplied USB cable IF-A28 to the USB connector of the probe.
Connect the USB plug on the other side of the cable to the probe USB connector of the CA-DP40 data processor.
Connection via a USB cable (using IF-A28)
Measurement start Measurement end
Input to the BNC plug
3.3 V CMOS input level (input up to 5 V)
Input trigger and external
synchronization signals
BNC connector
Mini DIN plug
IF-A30/31/32
Mini B plug
IF-A28
t > 5 μsec
Mini DIN plug
USB plug
18
t > 5 μsec
Measurement start Measurement end
Data Processor