Konica Minolta CA-2500 Series Instruction Manual
Konica Minolta CA-2500 Series Instruction Manual

Konica Minolta CA-2500 Series Instruction Manual

2d color analyzer

Advertisement

Quick Links

2D Color Analyzer CA-2500 Series
Instruction Manual
Before using this instrument,
please read this manual.

Advertisement

Table of Contents
loading

Summary of Contents for Konica Minolta CA-2500 Series

  • Page 1 2D Color Analyzer CA-2500 Series Instruction Manual Before using this instrument, please read this manual.
  • Page 2 Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as a result of incorrect use of the instrument. Denotes a sentence regarding safety warning or precaution. Read the sentence carefully to ensure safe and correct use. Denotes a prohibited operation.
  • Page 3: Safety Warnings And Cautions

    Safety Warnings and Cautions To ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises. Warning (Failure to adhere to the following points may result in death or serious injury.) Do not use this instrument in places where flammable or combustible gases (gasoline etc.) are present. Doing so may cause fire. Always use the AC adapter and power cord supplied as a standard accessory or optional (AC-A312), and connect it to indoor AC outlet of rated voltage 100-240 V 50/60 Hz. Failure to follow either of these may result in damage to instrument or AC adapter, fire or electric shock. Insert the power plug fully and securely. If the power plug is not pushed complete inserting may cause fire or electric shock. If this instrument is not used for a long time, disconnect AC adapter from AC outlet. Accumulated dirt or water on prongs of AC adapter plug may cause fire. Be sure to remove before use.
  • Page 4: Introduction

    Introduction Thank you for purchasing our 2D Color Analyzer CA-2500 series. This instrument measures unevenness of luminance and color on various display unit or projector, backlight and car instrument panel in 2 dimension. Read this manual carefully before use. "CA-2500" and CA-2500S are applied as representative examples for explanation and illustration respectively in this instruction manual. Only those with different specification are indicated model by model. Packaging material Retain accompanying packaging materials (carton, protector, and plastic bag) for future use. This is delicate measurement instrument. Use packaging materials supplied in purchasing in case this instrument needs to be transferred for such purpose as maintenance in Konica Minolta-authorized service facility. These packaging materials are useful for minimizing shock or vibration to this instrument in such situation.Should any of these packaging materials be lost or broken, contact the nearest Konica Minolta-authorized service facility.
  • Page 5: Lens And Filter Inside Lens Mount On Body

    Lens and filter inside lens mount on body l Be sure that there is no dirt on surface of lens and filter inside lens mount on body before measurement. Correct measurement result cannot be obtained if dirt, dust or handsoil is left or uncleaned. l Do not touch surface of lens and filter inside lens mount on body with hand. l Lens and filter inside lens mount on body could become fogged up if ambient temperature fluctuates under high humidity.
  • Page 6: Cleaning

    Cleaning Body l If this unit becomes dirty, wipe with dry and soft cloth. Do not use organic solvent like benzine or thinner and other chemical agent for cleaning. Should none of these methods be helpful, contact the nearest Konica Minolta-authorized service facility. Lens and filter inside lens mount on body l Should it be gotten dirt or dust, wipe off with dry and soft cloth or lens cleaning paper. Do not use organic solvent like benzine or thinner and other chemical agent for cleaning. Should none of these methods be helpful, contact the nearest Konica Minolta-authorized service facility. Notes on transfer l Use the original packaging materials to minimize the shock and vibration to the product when transporting it. l Pack the product and all of its accessories together in one package before sending them to us for repair, inspection and/or calibration.
  • Page 7: Table Of Contents

    CONTENTS Safety Warnings and Cautions ………………… 1 Setting measurement distance …………………23 Caution ……………………………………………………… 24 Introduction ………………………………………… 2 Example of reference position to read distance …25 Note on use ………………………………………… 2 How to read distance in case of standard lens or telephoto lens … 25 Operating environment …………………………………… 2 How to read distance in case of wide lens ……………… 25 This instrument ……………………………………………...
  • Page 8: Standard Accessory

    Standard Accessory Standard and optional accessories are available with the instrument. l The shape of some products may be different from those shown. Attached lens cap, mount cap, lens hood and hood cap corresponding with lens type for each model. Lens Cap l Place on lens for protection in not using this instrument. For standard or telephoto lens For wide lens For standard or For wide lens telephoto lens Mount Cap l Place on lens mount for protection when lens is removed from body.
  • Page 9: Optional Accessory

    Data Management Software CA-S25w l For data management of this instrument from PC. USB Cable (3 m) IF-A18 l Used for communication between this instrument and PC. Calibration Data DVD l Supplied with each lens. Optional Accessory Soft Case CA-A60 l Used to keep this instrument and accessories or carry them with hand. Never use for transfer. Tripod CS-A3 Pan Head CS-A4 Mura Evaluation Software CA-Mura l This program is an add-on software for CA-S25w.
  • Page 10: System Configuration

    System Configuration CA-2500A (With all lenses) Wide lens CA-A58 Hood cap Data management Mura Evaluation Lens hood (For wide) Mount cover (For wide) software software CA-A75 CA-S25w CA-Mura CA-A69 CA-2500 (on the market) Standard lens CA-A57 Hood cap Lens hood (For standard/telephoto, macro) USB cable (For standard/...
  • Page 11 CA-2500SW (With standard and wide lenses) Wide lens CA-A58 Lens hood Hood cap (For wide) Data management Mura Evaluation (For wide) Mount cover CA-A75 software software CA-A69 CA-S25w CA-Mura CA-2500 (on the market) Standard lens CA-A57 Hood cap Lens hood (For standard/telephoto, macro) USB cable (For standard/...
  • Page 12: Names And Functions Of Parts

    Names and Functions of Parts <Body> Handle Lens attaching portion Lens mount Air duct Power switch USB connector AC adapter input terminal Screw hole for fixing (ISO screw) Screw hole for fixing (Tripod screw) ○ Power switch Switches this instrument on/off. [|] side for ON; [ ] side for OFF. Green LED lights up while switched on. ………………………………………… (p.19) AC adapter input terminal To which accessory AC adapter is connected.
  • Page 13: Lens

    <Lens> Focus ring Distance indicator Lens hood Lens Body and lens arranged in a pair are calibrated. Pair those with identical numbers. CA-2500A, CA-2500S, CA-2500SW, CA-2500ST Standard lens CA-2500A, CA-2500W, CA-2500SW, CA-2500WT Aperture ring Wide lens CA-2500A, CA-2500T, CA-2500ST, CA-2500WT (Low magnification) (High magnification) Telephoto lens Macro ring Lens hood Protects lens. Remove this once when replacing lens. Focus ring Adjusts focus.
  • Page 15: Installing

    Installing...
  • Page 16: Goods/Environment To Prepare On Customer's Side

    Goods/Environment to Prepare on Customer's Side Get prepared for below goods/environment before using this instrument. Controlling PC : Windows 10 Pro 32 bit, 64 bit ® Windows 11 Pro ® (English, Japanese, or Chinese (simplified font) version of the above OS) The PC must meet the system requirements specified by the applicable OS or listed below, whichever is higher. CPU : Equivalent to Pentium 4, 2.8 GHz or better ® Memory : 1024 MB or more Hard disk drive : At least 80 MB of available disk space is required on the system drive (drive where the OS is installed). To install the calibration data, the following disk space is required for each lens: • For a standard lens: approximately 540 MB • For a wide angle lens: approximately 470 MB...
  • Page 17: Installing

    Installing Warning for safety Do not use this instrument in places where flammable or combustible gases (gasoline etc.) are present. Doing so may cause fire. Caution for safety Do not place this instrument on unstable or sloping surface which may drop or overturn it. Dropping or overturning may injure someone around. Take care not to drop this instrument when carrying. To set this instrument on tripod or jig, use screw hole for fixing at the bottom of this instrument. 2 type holes are available as follows; Tripod screw hole: To set on tripod. Screw depth is 6.5 mm. ISO screw hole: To set on jig. User ISO 5 mm screw and screw depth is 6.5 mm. 186.2 68.2 ISO screw hole Optical axis Tripod screw hole UNC 1/4 Depth 6.5...
  • Page 18: Placing Lens

    Placing Lens Below shows lens type accompanying respective CA-2500 series. Lens hood is served as standard corresponding to lens type. CA-2500A Standard, wide and telephoto lenses, and 2 macro rings CA-2500S Standard lens CA-2500W Wide lens CA-2500T Telephoto lens and 2 macro rings CA-2500SW Standard and wide lenses CA-2500ST Standard and telephoto lenses, and 2 macro rings CA-2500WT Wide and telephoto lenses, and 2 macro rings Operation procedure Remove mount cover from lens mount on body.
  • Page 19: Replacing Lens

    l Check whether body and lens have identical serial numbers. Serial number can be found on label of body and lens side respectively. 警告: カバーを開けないでください。 To pursue high-accuracy measurement, calibration was done on maker's WARNING: Do not remove covers. WARNUNG: Niemals das gehause offnen. side for each unit before shipment combining with each accessory lens 1 XXXXXX for CA-2500 series. This means that attached calibration coefficient data to each unit is valid exclusively for the packaged combination of unit and lens. l Set lens type to use in software CA-S25w. ( For detail, see the instruction manual for data management software CA-S25w.) This enables the application of appropriate calibration coefficient data depending on usage condition which would finally give correct measurement result.
  • Page 20: Connecting Ac Adapter

    Connecting AC Adapter Use accessory AC adapter for power source. Warning (Failure to adhere to the following points may result in death or serious injury.) Always use the AC adapter and power cord supplied as a standard accessory or optional (AC-A312), and connect it to indoor AC outlet of rated voltage 100-240 V 50/60 Hz. Failure to follow either of these may result in damage to unit, fire or electric shock. If this instrument is not used for a long time, disconnect AC adapter from AC outlet. Accumulated dirt or water on prongs of AC adapter plug may cause fire and should be removed. Be sure to remove before use. Do not forcibly pull any part on power cord when unplugging since this may cause fire or electric shock. Gently disconnect by holding plug. Also, do not handle power plug with wet hands. Doing so may cause electric shock.
  • Page 21: On ( | ) / Off ( ) Of Power Switch

    ON ( | ) / OFF ( ) of Power Switch To ON Press [ | ] side of power switch to To OFF Press [ ] side of power switch to OFF after completion of measurement. Connecting with PC Before connecting CA-2500 body with PC, install software CA-S25w in PC. (For detail, see the Installation Guide for data management software CA-S25w.) Use accessory USB cable (3 m) IF-A18 for this purpose.
  • Page 23: Measurement

    Measurement...
  • Page 24: Starting Up And Ending System

    Starting Up and Ending System Starting up Switch power on as instructed. If not, erroneous operation could occur. Switch power on for body. (Press [ | ] side.) l Motor sounds indicating controller inside machine starts initialization. l Fan inside body starts to run. l 1 5-minute warm-up at least (to leave in this state after switching on.) is recommended for accurate measurement. Start up Windows by turning on controlling PC. Connect PC and CA-2500 with USB cable. l P C recognizes that CA-2500 has been connected.
  • Page 25: Setting The Measurement Distance And The Pint Ring Range Index

    Setting the measurement distance and the pint ring range index Setting measurement distance Five types of objective lenses are available for CA-2500 depending on the size of measuring objects. The relation between the size of measuring objects and the measurement distance for each lens is as follows: CA-2500A, CA-2500S, CA-2500SW, CA-2500ST CA-2500A, CA-2500W, CA-2500SW, CA-2500WT Standard lens Wide lens Size of measurement object Size of measurement object Measurement distance Measurement distance (One-side length of square) (One-side length of square) 98 mm approx.
  • Page 26: Caution

    CA-2500A, CA-2500S, CA-2500SW, CA-2500ST Wide lens Distance With distance label On indicator ○ 0.2 m Indicator position (See Figure 3-1.) ○ 0.24 m Indicator position ○ 0.3 m Indicator position (See Figure 3-2.) ○ 0.5 m Indicator position ○ Indicator position ○ Indicator position CA-2500A, CA-2500S, CA-2500SW, CA-2500ST Wide lens Distance With distance label On indicator ○...
  • Page 27: Example Of Reference Position To Read Distance

    Example of reference position to read distance How to read distance in case of standard lens or telephoto lens The ring (focus ring) on which number values are written rotates in case of standard lens or telephoto lens. If the "distance indicator" includes decimal point, "." is determined as reference position; and if it does not include decimal point, the center of the number or [ ] is determined as reference position. Standard lens Figure 1-1 Figure 1-2 Figure 1-3 Example when setting to "0.25" Example when setting to "diverting Example when setting to "diverting point between 0.25 and 0.3 in 1:1"...
  • Page 29: Others

    Others...
  • Page 30: Outer Dimensions

    Outer Dimensions Reference position 72.7 52.8 警告: カバーを開けないでください。 WARNING: Do not remove covers. WARNUNG: Niemals das gehause offnen. 1 XXXXXX 192.1 186.2 61.2 128.9 68.2 Rubber foot diameter : Ø15 M5, Depth : Ø6.5 Tripod socket 81.2 UNC 1/4, Depth : 6.5 95.2 Positioning hole Depth : 4.5 111.2 Ø5 positioning hole Depth : 4.5 125.2...
  • Page 31: Error Check

    Error Check Should error be found in this instrument, try corrective actions shown in the following table. If this does not help, this instrument has possibly been broken. Please contact the nearest Konica Minolta-authorized service facility with indicated error number. Error Symptom Item to Check Corrective Action Power does not turn Has AC adapter been plugged in Connect AC adapter. outlet? Does not normally start up. Has AC adapter been connected Connect AC adapter.
  • Page 32: Main Specification

    Main Specification Model CA-2500S CA-2500W CA-2500T CCD image sensor (monochrome); 2/3-inch; Effective number of pixels: 1,000 x 1,000 pixels; Light receptor Equipped with XYZ filter (closely matches CIE 1931 color-matching function) and ND filter Interchangeable Standard, wide, and telephoto lenses; low-magnification and high-magnification Lens macro rings (for use with telephoto lens) Measurement points (Resolution) 980 x 980 (490 x 490 or 196 x 196 selectable by using Data Management Software CA-S25w) Color indication modes XYZ, L v xy, L v u’v’, T uv, Dominant wavelength, Excitation purity, L v contrast Display modes Pseudocolor, Chromaticity diagram, Spot, Cross section, Color difference Standard lens Wide lens...
  • Page 33 *1: Error in angle of view: 7% *2: Measurement time differs depending on brightness of measurement object, PC performance, and data processing contents. The specifications above were obtained under Konica Minolta's measurement conditions shown below: PC CPU : Intel Core™ i7-3770 3.40 GHz ® Memory : 8 GB Data processing : Pseudocolor display of luminance data Resolution : 980 x 980 Shutter speed : Y measurement : 1/64 sec., XZ measurement : 1/32 sec. Measurement subject brightness: Standard/wide lens: Approx. 80 cd/m , Telephoto lens: Approx. 300 cd/m Low-magnification macro ring and telephoto lens: Approx. 400 cd/m High-magnification macro ring and telephoto lens: Approx. 600 cd/m * The measurement time becomes longer when the object is dark. The longest measurement time is approx. 14 seconds with 1-time integration, approx. 45 seconds with 4-time integration, approx. 3 minutes with 16-time integration, approx. 11 minutes with 64- time integration and approx. 44 minutes with 256-time integration *3: The specifications above were obtained under Konica Minolta's measurement conditions shown below: Measurement subject brightness: Standard/wide lens: Approx. 35 cd/m , Telephoto lens: Approx. 140 cd/ m Low-magnification macro ring and telephoto lens: Approx. 250 cd/m High-magnification macro ring and telephoto lens: Approx. 250 cd/m Distance : Minimum distance of each lens Evaluation : Based on the average obtained within 10% range at the center of the screen Temperature : 23°C±2°C...
  • Page 34 < CAUTION > KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE USE OF OR INABILITY TO USE THIS PRODUCT.
  • Page 35 © 2013-2023 KONICA MINOLTA, INC. 9222-A6DJ-11 CDEDDK...