Setting/Changing the Target

1� Target
A target is the value used as a reference to determine how much the measured value differs from it. Use one of the following
methods to set a target.

(1) By measurement and registration

(2) By entering values
The target is used in common for each measurement mode (Lv x y, Lv u'v', Lv Tcp duv, X Y Z, and Dominant wavelength).
By default, factory-set target values are preset in all channels.
2� By Measurement and Registration
Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key�
The setting screen is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[Target], and then press the [RIGHT] or
[ENTER] key�
3
Press the [UP] or [DOWN] key to select
[Measure and Register], and then press
the [RIGHT] or [ENTER] key�
4
The screen for selecting a probe is
displayed� Press the [UP] or [DOWN] key
to select the probe to register, and then
press the [RIGHT] or [ENTER] key�
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
MENU
M e a s   C o n d
U s e r   C a l i b
T a r g e t
O p t i o n
S y s t e m
82
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
E n t e r   V a l u e
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
E n t e r   V a l u e
Trigger mode 0-cal AUTO
M e a s u r e   a n d   R e g i s t e r
01
E n t e r   V a l u e
02
03
04
05
06
07
08
09
10