When the pop-up screen for single-point
calibration is displayed, press the
[ENTER] key to perform measurement�
When the measured value is displayed,
press the [ENTER] key to hold the value�
8
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to select the field of
a calibration value for user calibration,
and then press the [ENTER] key�
9
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to set a calibration
value, and then press the [ENTER] key�
Press the [ESC] key to cancel configuration of the
setting and restore the original value.
10
Repeat steps 8 and 9 to set the
calibration value, move to [Complete],
and then press the [ENTER] key to
confirm your entry�
(If the entered value is out of the available setting
range, the screen returns to the value entry screen.
Enter another value.)
11
Press the [ESC] key�
The measurement screen is displayed on the LCD
screen.
The specified values will be kept even if the POWER
switch is set to OFF ().
MENU
User Calib > Single-point calibration
Single-point calibration
Start measurement
01 CH01
Measured Value
Measured Value
Target Value
MENU
User Calib > Single-point calibration
Single-point calibration
Stop measurement
01 CH01
Measured Value
Measured Value
Target Value
MENU
User Calib > Single-point calibration
01 CH01
Measured Value
MENU
User Calib > Single-point calibration
Single-point calibration
Edit Value
To confirm by pressing the ENTER button
01 CH01
in the "Comp"
Measured Value
 160.2        0.3132       0.3257
Measured Value
Target Value
MENU
User Calib > Single-point calibration
01 CH01
Measured Value
 160.2        0.3132       0.3257            160.0       0.3130       0.3260
71
校正値
Lv                                              x                                                y
Meas
Complete
校正値
Lv                                              x                                                y
 160.2             0.3132            0.3157
Stop
Complete
Calibration Value
Complete
校正値
Lv                                              x                                                y
 160.2             0.3132            0.3257
Complete
Complete
Calibration Value
Complete