Konica Minolta CA-410 Instruction Manual page 75

Display color analyzer, probe+ data processor
Hide thumbs Also See for CA-410:
Table of Contents

Advertisement

7
The screen for selecting a color is
displayed� Press the [UP] or [DOWN] key
to select the color to measure and then
press the [ENTER] key�
8
When the pop-up screen for RGB+W
matrix calibration is displayed, press the
[ENTER] key to perform measurement�
When the measured value is displayed,
press the [ENTER] key to hold the value�
9
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to select the field of
a calibration value for user calibration,
and then press the [ENTER] key�
10
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to set a calibration
value, and then press the [ENTER] key�
Press the [ESC] key to cancel configuration of the
setting and restore the original value.
11
Repeat steps 9 and 10 to set the
calibration value, move to [Complete],
and then press the [ENTER] key to
confirm your entry�
12
Repeat steps 7 to 11 to set calibration
values for W, R, G, and B, move to
[Complete], and then press the [ENTER]
key to confirm your entry�
(If the entered value is out of the available setting
range, the screen returns to the value entry screen.
Enter another value.)
13
Press the [ESC] key�
The measurement screen is displayed on the LCD
screen.
The specified values will be kept even if the POWER
switch is set to OFF ().
MENU
User Calib > RGB+W matrix calibration
Select Color
01 CH01
Measured Value
MENU
User Calib > RGB+W matrix calibration
RGB+W matrix calibration
Select Color
White
Start measurement
01 CH01
Measured Value
Measured Value
Target Value
MENU
User Calib > RGB+W matrix calibration
RGB+W matrix calibration
Select Color
White
Stop measurement
01 CH01
Measured Value
Measured Value
Target Value
MENU
User Calib > RGB+W matrix calibration
RGB+W matrix calibration
Select Color
White
Edit Value
01 CH01
Measured Value
To confirm by pressing the ENTER button in the "Comp"
 160.2        0.3132       0.3257           
Measured Value
Target Value
MENU
User Calib > RGB+W matrix calibration
Select Color
01 CH01
Measured Value
 160.2        0.3132       0.3257            160.0       0.3130       0.3260
73
Calibration Value
Complete
校正値
Lv                                              x                                                y
Meas
Complete
校正値
Lv                                              x                                                y
 160.2             0.3132            0.3157
Stop
Complete
校正値
Lv                                              x                                                y
 160.2             0.3132            0.3157
Complete
Complete
Calibration Value
Complete