2.1.10

HAL 100 halogen illuminator

(1)
Attachment
illuminator
• Remove cover from transmitted-light tube
(2-11/3).
• Insert dovetail 2-12/2) of lamp housing (2-11/1)
in transmitted-light tube (2-11/3) on the
instrument rear and use the SW 3 ball-headed
screwdriver to tighten clamping screw (2-11/2).
• Connect 3-pin lamp plug (2-11/5) to 3-pin
12 V/100 W socket (2-11/4) on the instrument
rear.
(2)
Adjustment
illuminator
a)
Coarse adjustment
• Loosen clamping screw (2-11/2) and remove the
halogen illuminator from the microscope stand.
• Switch on microscope as described under 0.
• Direct the light beam against a projection area
(wall) approx. 3 m away.
Do not look into the light exit opening
of the illuminator.
• Set adjusting screw (2-12/3) using SW 3 ball-
headed screwdriver in such a way that both
images of the lamp coil are as sharply focused
as possible on the projection screen.
• Then set adjusting screws (2-12/4 and 5) in such
a way that the lamp coil of one image exactly
fills
the
gaps
(2-12/1).
B 40-075 e
10/03
of
HAL 100
halogen
of
HAL 100
halogen
in
the
reflector
START-UP
Initial start-up
Fig. 2-11
image
Fig. 2-12
Attachment of HAL 100 halogen
illuminator
Adjustment of HAL 100 halogen
illuminator
Carl Zeiss
2-9