Zeiss Axioskop 2 plus Operating Manual page 56

Routine microscope
Table of Contents

Advertisement

Carl Zeiss
Fig. 2-37
Setting ergo-drive
• Mount the additional sleeves back on the drive knobs in reverse order and tighten both the clamping
screws.
(5)
Setting the smoothness (torque) of the two drive knobs of the ergo-drive
The ergo-drive is set to a medium smoothness value at the factory. This setting can be changed as
follows.
a)
Adjusting the X direction
• Remove the additional sleeves (2-37/1 and 3) from the drive knobs if necessary. Loosen the clamping
screws to do so.
• Push the X drive knob (2-37/5) down and the Y drive knob (2-37/7) up.
• Hold the X drive knob (2-37/5) and turn the bright knurled ring (2-37/6) above it to the right (easy
action) or left (sluggish action) until the desired smoothness is achieved.
b)
Adjusting the Y direction
• Hold the Y drive knob (2-37/7) and turn the bright knurled sleeve (2-37/8) above it to the right
(sluggish action) or left (easy action) until the desired smoothness is achieved.
• Replace the additional sleeves if necessary and tighten the clamping screws.
2-24
START-UP
Attachments and conversions
(3)
On the mechanical stages with ergo-drive, the
travel range of the X and Y adjustment can be
extended by a max. 15 mm by axial movement of
the drive buttons.
(4)
Both drive buttons are fitted with additional
sleeves. These serve to provide an even finer
setting of the object position. They can be
removed if faster object movement is more
important.
• First loosen the two clamping screws (2-37/4) of
the bottom additional sleeve (2-37/3)
remove it downwards, then loosen both
clamping screws (2-37/2) of the top additional
sleeve (2-37/1) and pull this down and off as
well.
Adjusting the travel on ergo-drive
Removing and mounting additional
sleeves