Measurement Parameters - Nellcor OxiMax N 5500 Service Manual

Patient monitor
Table of Contents

Advertisement

9.7 Measurement Parameters

ECG
Measurement Range
Accuracy
Leads
Lead Off Detection
Input
Input Dynamic Range
Voltage Range
Signal Width
Output
Frequency Response (Bandwidth)
Low Extend
Filter
Monitor
ECG Size (sensitivity)
Display Sweep Speeds
Pacemaker Detection
Defibrillator Discharge
Recovery
Respiration rate
Technique
Range
Accuracy
Leads
Display Sweep Speeds
Lead Off Condition
Heart Rate
20 BPM to 250 BPM
±3 BPM or ±5%, whichever is greater
ECG (Electrocardiograph)
3 / 5 Lead, user selectable I, II, III, V, aVR, aVL, aVF
Detected and displayed
±5 mV AC, ±300 mV DC
±0.5 mV to ±5 mV
40ms to 120ms (Q to S)
ECG (Electrocardiograph)
0.05 Hz to 40 Hz
0.5 Hz to 30 Hz
0.5 Hz to 40 Hz
1.25, 1.7, 2.5, 5.0, 7.5, 10.0, 15.0, 20.0 (mm per 1mV)
12.5 mm/sec, 25.0 mm/sec, and 50.0 mm/sec
Indicator on waveform display (user selectable)
< 5 sec per IEC601-2-27
Trans-thoracic impedance
3 breaths/min to 120 breaths/min
±3 breaths/min
RA to LA
6.25 mm/s, 12.5 mm/s, 25.0 mm/s
Detected and displayed
N5500 Service Manual
Specification
75