Trends - Nellcor OxiMax N 5500 Service Manual

Patient monitor
Table of Contents

Advertisement

SpO
2
Range
Low Perfusion
Accuracy
Display Sweep Speeds
Range
Accuracy
1
Adult specification are shown for OxiMax MAX-A and MAX-N sensors with the N5500. Neonate
specifications are shown for OxiMax MAX-N sensors with the N5500. Saturation accuracy will vary
by sensor type.
2
Applicability(accuracy with motion): OxiMax sensors.
3
Specification applies to monitor performance and was validated with Biotek and Nellcor simulators
Temperature
Probe Type
Parameter displayed
Range
Display Accuracy
Probe Accuracy

9.8 Trends

Types
Memory Storage
Graphical Format
Display Range
Tabular Format
Display Time Interval
%Saturation
1% to 100%
0.03% to 20%
Without Motion-Adults
Without Motion-Neonate
With Motion
-Adults/Neonate
Low Perfusion
12.5 mm/sec, 25.0 mm/sec, and 50.0 mm/sec
Pulse Rate
20 BPM to 250 BPM
Without Motion
With Motion
Low Perfusion
Thermistor probe (Monotherm™ temperature probes)
T1, T2
15° C to 45° C
±0.1° C (25° C to 45° C)
±0.2° C (15° C to less than 25° C)
±0.1° C
Graphical and Tabular
24 hours, nonvolatile, 20 seconds data interval
One graph for all parameters
Also user-selectable each parameter to be desired
2 hours
One table for all variables.
18 rows (date, parameter labels, and 16 variable rows)
20 sec, 1 min, 2 min, 3 min, 5 min, 10 min, 15 min, 20 min
N5500 Service Manual
1
70% to 100% ±2 digits
1
70% to 100% ±3 digits
1% to 69% unspecified
2
70% to 100% ±3 digits
3
70% to 100% ±2 digits
1% to 69% unspecified
3
20 BPM to 250 BPM ±3 digits
normal physiologic range
(55 to 125 BPM) ±5 digits
3
20 BPM to 250 BPM ±3 digits
Specification
77