Perform A D3 And D4 Density Calibration (T-Series Sensors Only) - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
9.9
Perform a D3 and D4 density calibration (T-
Series sensors only)
For T-Series sensors, the optional D3 and D4 calibration could improve the accuracy of the
density measurement if the density of your process fluid is less than 0.8 g/cm
than 1.2 g/cm
If you perform the D3 and D4 calibration, note the following:
Configuration and Use Manual
3
.
Do not perform the D1 and D2 calibration.
Perform the D3 calibration if you have one calibrated fluid.
Measurement support
3
or greater
161