9.5 Perform a D3 and D4 density calibration (T-Series
sensors only)
For T-Series sensors, the optional D3 and D4 calibration could improve the accuracy of the density
measurement if the density of your process fluid is less than 0.8 g/cm
If you perform the D3 and D4 calibration, note the following:
Do not perform the D1 and D2 calibration.
Perform the D3 calibration if you have one calibrated fluid.
Perform both the D3 and D4 calibrations if you have two calibrated fluids (other than air and water). The
calibrations must be performed without interruption, in the order shown. Make sure that you are prepared
to complete the process without interruption.